欢迎来到匠仁医疗设备有限公司网站!

更多 > 新闻资讯
公司新闻
行业知识
检测答疑
图标 首页 >> 新闻资讯 > 行业知识

行业知识

内窥镜图像处理系统软镜主机维修

来源:http://www.jiangrenyiliao.cn/ 发布时间:2024-03-25 浏览量:0

内窥镜系统及周边设备维修范围:

Scope of maintenance for endoscope system and peripheral equipment:

软镜主机、软镜光源、超声乳化手柄、电子镜、纤维镜、摄像系统、摄像头、气腹机

Soft mirror host, soft mirror light source, ultrasonic emulsification handle, electronic mirror, fiber mirror, camera system, camera, pneumoperitoneum machine

高频电刀、冷光源、超声刀、动力系统、导光束、手术器械等

High frequency electric knife, cold light source, ultrasonic knife, power system, guided beam, surgical instruments, etc

医疗用内窥镜主要性能参数

Main performance parameters of medical endoscopes

(4)机械性能:

(4) Mechanical performance:

①密封性:内窥镜的密封性关系到成像质量,如密封不好就容易渗水,破坏光学系统,影响观察。通水阀镜鞘与膀胱镜的锥体配合处应密合,在1min内渗水应不超过5滴,否则配合不好容易漏水影响手术。

① Sealing: The sealing of the endoscope is related to the imaging quality. If the sealing is not good, it is easy to leak water, damage the optical system, and affect observation. The fitting between the sheath of the water valve and the cone of the cystoscope should be tight, and the water leakage should not exceed 5 drops within 1 minute. Otherwise, poor fitting may cause water leakage and affect the surgery.

②表面粗糙度

② Surface roughness

③连接部位牢固性

③ Durability of connection parts

(5)绝缘性能:

(5) Insulation performance:

主要是电子内窥镜,相关电气隔离部位的绝缘结构,如与CCD摄像头相配接的目镜罩、冷光源接口、导光索等相关附件的绝缘性能。

Mainly the insulation structure of electronic endoscopes and related electrical isolation parts, such as the insulation performance of eyepiece covers, cold light source interfaces, light guide cables, and other related accessories that are matched with CCD cameras.

(6)绝热性能:

(6) Insulation performance:

在医用内窥镜中,由于内窥镜是侵入性检查工具,为了避免内窥镜工作的时候对人体的伤害,现在一般都采用冷光源,在光输出口设置红外滤光片,。大限度地限制红外光的输出。

In medical endoscopes, as endoscopes are invasive examination tools, in order to avoid harm to the human body during the operation of endoscopes, cold light sources are generally used, and infrared filters are installed at the light output outlet. Maximize the output of infrared light.

内窥镜测漏流程

Endoscopic leak detection process

1、内镜盖好防水帽挂于主机上;

1. Cover the endoscope with a waterproof cap and hang it on the host;

2、打开光源开关;

2. Turn on the light source switch;

3、将测漏器插入主机;

3. Insert the leak detector into the host;

4、检查测漏器压力;

4. Check the pressure of the leak detector;

20200403031114827

5、测漏器与镜子相连接;

5. The leak detector is connected to the mirror;

6、充气3秒钟看镜子前头是否漏气;

6. Inflate for 3 seconds to check if there is any air leakage in the front of the mirror;

7、将镜子完全放入水中;

7. Completely immerse the mirror in water;

8、使各旋钮均处于放松状态并旋转大小旋钮,检查有无漏气(约30秒);

8. Place all knobs in a relaxed state and rotate the large and small knobs to check for air leakage (approximately 30 seconds);

9、将镜子取出擦干;

9. Take out the mirror and dry it;

10、测漏器与主机分离(约30秒,目的是放掉镜内的气体);

10. The leak detector is separated from the host (for about 30 seconds, with the aim of releasing the gas inside the mirror);

11、测漏器与镜子分离,操作完毕。

11. The leak detector is separated from the mirror and the operation is complete.

奥林巴斯纤维细镜先端组件的组装过程

The assembly process of Olympus fiber mirror tip components

二、清理先端头

2、 Clean the leading end

1.用抬钳座夹住先端头,盖玻璃面朝上,将1mm直径的钢棒对着钳子管出口位金属接头,用小锤敲击,直到把金属接头连着部分钳子管往下脱出,接着用小号一字批将残余在钳子管安装孔里的AB胶清理干净;

1. Use the pliers holder to clamp the front end, with the glass cover facing upwards. Place a 1mm diameter steel rod towards the metal joint at the outlet of the pliers tube, and tap it with a small hammer until the metal joint is partially detached from the pliers tube. Then, use a small batch to clean the remaining AB glue in the installation hole of the pliers tube;

2.将合适的钢棒插进空心的像束金属部,用斜口钳夹住露出先端头外的金属部,把金属头完全夹取出来。

2. Insert a suitable steel rod into the hollow bundle of metal parts, clamp the metal part exposed outside the tip with diagonal pliers, and completely remove the metal head.

3.用磨尖的小号批,在放大镜下清理像束安装孔壁的AB胶,(小心不能碰到物镜,以免刮花物镜)直到干净后把物镜倒出来,(有时需要用力敲击先端头才能倒出来)然后一手用镊子夹住物镜,一手用镊子夹镜头纸,沾酒精把贴在物镜上的光学胶拭擦掉,直镜片干净无尘。

3. Use a sharpened small batch to clean the AB adhesive on the wall of the image beam installation hole under a magnifying glass (be careful not to touch the objective lens to avoid scratching the objective lens) until it is clean. Then, pour out the objective lens (sometimes requiring strong tapping on the front end to pour it out), and use tweezers to clamp the objective lens with one hand, and use tweezers to clamp lens paper with the other hand. Wipe off the optical adhesive attached to the objective lens with alcohol until the lens is clean and dust-free.

4.用合适转头的手动转把光束安装孔里的光纤清理出来,当转到接近底的时候放慢以免磨花光束盖玻璃。

4. Use a suitable manual rotation head to clean the optical fibers from the installation hole of the beam. When the rotation is close to the bottom, slow down to avoid grinding the glass of the beam cover.

5.用镜头纸沾酒精清理像束及光束安装孔。

5. Clean the image beam and beam mounting holes with lens paper dipped in alcohol.

本文由内窥镜设备维修提供技术支持,更多的详细精彩内容请点击我们的网http://www.jiangrenyiliao.cn我们将会全心全意为您提供满意的服务。

This article provides technical support for endoscopic equipment maintenance. For more detailed and exciting content, please click on our website http://www.jiangrenyiliao.cn We will wholeheartedly provide you with satisfactory service.

s
新闻资讯
联系我们
  • 公司公司:匠仁医疗设备有限公司
  • 匠仁医疗山东济南维修部
  • 联系人:樊经理
  • 电话:13153199508
  • 地址:山东省济南市槐荫区美里东路3000号德迈国际信息产业园6号楼101-2室
  • 匠仁医疗湖南长沙维修部
  • 联系人:李经理
  • 电话:13873135765
  • 地址:湖南省长沙市雨花区劳动东路820号恒大绿洲14栋2409室
新闻资讯

樊经理:13153199508     李经理:13873135765

公司地址:山东省济南市槐荫区美里东路3000号德迈国际信息产业园6号楼101-2室  湖南省长沙市雨花区劳动东路820号恒大绿洲14栋2409室

     

鲁公网安备 37010402441281号