内窥镜常见故障怎么处理

来源:http://www.jiangrenyiliao.cn/ 发布时间:2022-06-30 浏览量:0

随着新医改政策的不断深化,医疗保险制度、医疗服务体系的不断完善,以及分级诊疗制度的试点与推行,人民群众医疗需求将不断提高,必将推动医疗器械消费的持续增加。在市场需求的刺激和经济持续稳定增长的背景下,的医疗器械产业还将处于快速发展期。
With the deepening of the new medical reform policy, the continuous improvement of the medical insurance system and the medical service system, as well as the pilot and implementation of the graded diagnosis and treatment system, the medical demand of the people will continue to increase, which will certainly promote the continuous increase of the consumption of medical devices. Under the background of market demand stimulation and sustained and stable economic growth, China's medical device industry will still be in a period of rapid development.
内窥镜设备维修已经成为了必备设备,很多地方的检查都用于内窥镜,可用于检查也可用于,宠物医院也用的较多,时间久了也都会大大小小的出现问题,很多小问题其实自己也都可以处理,比如常见侧漏、更换橡皮等,本公司都会将这些常见故障问题做分析、探讨、交流学习。
The maintenance of endoscope equipment has become a necessary equipment. In many places, the inspection is used for endoscopes, which can be used for inspection and treatment. Pet hospitals also use it more. Over time, there will be large and small problems. In fact, many small problems can be solved by ourselves, such as common side leakage, rubber replacement, etc. our company will analyze, discuss, exchange and learn about these common problems.
先端弯曲部的故障
Failure of the tip bend
一、镜面出现了缺口、破裂及磨损
1、 The mirror is chipped, cracked and worn
原因分析:
Cause analysis:
1、镜面缺口、破裂的主要原因是内镜先端部与周围坚硬物体发生碰撞,货先端部坠落地板,导致镜面玻璃的破碎。
1. The main reason for the notch and fracture of the mirror is that the tip of the endoscope collides with the surrounding hard objects, and the tip of the cargo falls to the floor, resulting in the breakage of the mirror glass.
2、内镜在清洗内清洗时,先端部与清洗槽多次接触造成磨损。
2. When the endoscope is cleaned in the cleaning and disinfection chamber, the tip contacts with the cleaning tank for many times, resulting in wear.
内窥镜设备维修
预防方法:
Prevention methods:
1、台车侧面增加海绵垫及防水胶垫,减少与台车的碰撞。
1. Sponge pad and waterproof rubber pad are added at the side of the trolley to reduce the collision with the trolley.
2、在清洗槽增加保护垫,避免碰撞及磨损。
2. Add a protective pad in the cleaning tank to avoid collision and wear.
二、镜面模糊不清
2、 Blurred mirror
原因分析:
Cause analysis:
物镜表面被清洗残留污垢物覆盖
The surface of the objective lens is covered by residual dirt from cleaning and disinfection
预防方法:
Prevention methods:
用纱布蘸酒精反复清洗镜面。
Wash the mirror repeatedly with gauze dipped in alcohol.
注意:使用纱布清洁的时候需要注意水气喷嘴方向,以免纱布堵塞水气喷嘴。
Note: when cleaning with gauze, pay attention to the direction of the water vapor nozzle to prevent the gauze from blocking the water vapor nozzle.
三、喷嘴堵塞,不能正常送水及送气
3、 Nozzle blocked, unable to supply water and air normally
原因分析:
Cause analysis:
在内镜使用中,由于腔内充气膨胀,造成内镜管道内气压低,以及毛细血管作用,导致粘液或血液等进入水气通道。加上水气管道狭小,不易清洁,长期积累导致喷嘴堵塞。
In the use of endoscopes, due to the inflation and expansion in the cavity, the pressure in the endoscope pipe is low, and the capillary effect causes mucus or blood to enter the water gas channel. In addition, the water and gas pipeline is narrow and difficult to clean, which leads to nozzle blockage due to long-term accumulation.
预防方法:
Prevention methods:
1、加强床旁清洗,从病人体内取出内镜后,先反复送水、送气冲洗管道。
1. Strengthen the bedside cleaning. After taking out the endoscope from the patient's body, repeatedly send water and air to wash the pipeline.
2、使用纱布清洗先端头时,不要正对喷嘴擦拭。
2. When cleaning the tip with gauze, do not wipe it directly against the nozzle.
3、每次清洗使用水枪或者50ml注射器注水清洗。
3. Use a water gun or a 50ml syringe to inject water for each cleaning.
以上便是问题的相关解答内容了,如果您想获取更多相关信息,就来关注我们网站http://www.jiangrenyiliao.cn吧!
The above is the relevant answers to the questions. If you want to get more relevant information, please follow our website http://www.jiangrenyiliao.cn Come on!
s
新闻资讯
联系我们
  • 公司公司:匠仁医疗设备有限公司
  • 匠仁医疗山东济南维修部
  • 联系人:樊经理
  • 电话:13153199508
  • 地址:山东省济南市槐荫区美里东路3000号德迈国际信息产业园6号楼101-2室
  • 匠仁医疗湖南长沙维修部
  • 联系人:李经理
  • 电话:13873135765
  • 地址:湖南省长沙市雨花区劳动东路820号恒大绿洲14栋2409室
新闻资讯

樊经理:13153199508     李经理:13873135765

公司地址:山东省济南市槐荫区美里东路3000号德迈国际信息产业园6号楼101-2室  湖南省长沙市雨花区劳动东路820号恒大绿洲14栋2409室

     

鲁公网安备 37010402441281号